Inductively Coupled Plasma Mass Spectrometer (ICPMS)
  • Overview
  • Quick View Leaflet
  • Detail Technical Brochure
  • Application Procedure
Inductively Coupled Plasma Mass Spectrometer (ICPMS)
Overview

Inductively Coupled Plasma Mass Spectrometer (ICPMS) is an import substitute and a versatile analytical instrument capable of trace / ultra-trace / isotope ratio analysis of elements in aqueous samples. It is a compact, reliable and highly sensitive analytical tool with 20x106 counts per second per part per million part of Indium and detection limit is parts per billion to parts per trillion for most of the elements. ICPMS finds numerous applications in various fields like Nuclear, Environment, Geological, Forensic; industries like Semi-conductor, Chemical/ Pharmaceutical, Petro-chemical and Research & Development institutes.

Inductively Coupled Plasma Mass Spectrometer (ICPMS)
Quick View Leaflet

Inductively Coupled Plasma Mass Spectrometer (ICPMS), an import substitute, is a very versatile analytical instrument capable of trace and ultra-trace analysis and isotope ratio analysis in aqueous samples.

 SALIENT FEATURES OF THE ICPMS

  • ICPMS is a compact, reliable and high sensitivity analytical tool consisting of an inductively coupled plasma ion source, quadrupole mass analyzer, rotary & turbomolecular pump for clean vacuum, easy sample introduction, channel electron multiplier detector and PC based interface & control.

  • Its sensitivity is 20x106 counts per second per 1 ppm of Indium and detection limits are in the range of ppt to ppb for most of the elements across the periodic table.

 APPLICATIONS

  • ICPMS finds numerous applications in various fields like Nuclear, Environment, Geological, Forensic; industries like Semi-conductor, Chemical/ Pharmaceutical, Petro-chemical and Research & Development institutes.


Inductively Coupled Plasma Mass Spectrometer (ICPMS)

 

Detail Technical Brochure

Inductively Coupled Plasma Mass Spectrometer (ICPMS) is a very versatile analytical instrument capable of trace and ultra-trace analysis and isotope ratio analysis in aqueous samples. It consists of an inductively coupled plasma ion source, quadrupole mass analyzer, rotary & turbomolecular pump for clean vacuum, easy sample introduction, PC based interface & control & Channel electron multiplier detector. Its sensitivity is 20x106 counts per second per 1 ppm of Indium and detection limits are ppt to ppb for most of the elements.

APPLICATIONS

  • ICPMS finds numerous applications in Nuclear field, Environmental fields, Geological laboratories, Semi-conductor industries, Chemical/ Pharmaceutical industries, Petro-chemical industries, Forensic laboratories and Research & Development institutes.

SPECIFICATIONS

a. Ion source

  • Source
  • : Inductively coupled plasma
  • RF Power
  • : Upto 2kW @ 27.12MHz
  • Sample introduction system
  • : Aqueous samples through nebulizer & spray chamber
  • Ion Optics
  • : Double deflection type with Einzel lens

    b. Quadrupole Mass Analyzer

  • Rod diameter
  • : 16 mm
  • Rod length
  • : 250 mm
  • Pre-filter/post filter
  • : 30 mm
  • RF frequency
  • : 1.2 MHz
  • Resolution
  • : 1 amu (adjustable)
  • Mass range
  • : 1-300 amu
  • Sensitivity
  • : 10 x 106 counts /sec for 1 parts per million of Indium

    c. Vacuum System

  • Vacuum system
  • : Two turbo molecular pumps (600 l/s & 300l/s)
  • Roughing pump
  • : Two rotary pumps (35 m3/h & 10 m3/h)

    INFRASTRUCTURE REQUIRED

    a. Mechanical Fabrication Workshop with the following facilities

    • Conventional Lathe Machine
    • CNC Lathe Machine
    • Centreless Grinding Machine
    • Vertical Milling Machine
    • TIG Welding
    • Electropolishing
    • CNC jig Boring

    b. Facilities for Inspection of mechanical sub - assemblies

    • Normal Precision Measuring Instruments
    • 3D-Co-ordinate Measuring Machine (L.C.—1 micron)
    • MSLD ( Helium Leak Detector ) for Vacuum Testing (Leak Rate: 10-12 Torr Litre/sec)

    c. Electronics Assembly & Test Facilities

    • RF load, RF power meter
    • PCB assembly, SMD workbench & Instrument rack wiring facility
    • Normal measuring instruments like multimeters & 100MHz Oscilloscopes
    • Function generator ( 1MHz )
    • 6 & 1/2 Digit Digital Voltmeter or Digital Multimeter
    • R.F. Digital Voltmeter with attenuator
    • High Voltage measurement system

    d. Space

    • 300 square feet space is required for assembly, testing & installation of the Inductively coupled plasma mass spectrometer. Venting arrangement is required for Ozone produced by the from plasma

    e. Power Supply

    • 230 Volts, 32 Ampere, Three Phase, 4kVA with RF ground

    MATERIAL AVAILABILITY

    • Most of the components are available within the country and some components need to be imported through local agents.

    MANPOWER

    • A physics post graduate, electronics engineer, mechanical engineer & 2 trained technicians will be required for the development of this instrument.

    Who Should apply


    Interested parties with Engineering & Scientific knowledge, good financial background and adequate experience of products manufacturing & fabrication with technical capability in the area of interested technology and having or interested in setting –up facilities for production would be preferred.


    How to apply

    Send your Technology Transfer Application form duly filled and signed alongwith a Demand Draft/Bankers cheque of Rs. 500/- (for Indian entities) or US $50/- (for foreign entities) drawn in favour of “Accounts Officer, BARC” as application processing fee on following address: 

    Head, Technology Transfer and Collaboration Division 
    Bhabha Atomic Research Centre,
    Trombay, Mumbai – 400 085, India.

    Fax : +91-22-25505151


    Note:Applications without processing fee as applicable above of Rs. 500/- or US $50/- for each technology will not be considered.


    Click here for Technology Transfer Application Form

    Contact Us

    Head, 
    Technology Transfer and Collaboration Division
    Bhabha Atomic Research Centre,
    Trombay, Mumbai 400 085.

    Fax : 091-022- 25505151
    Email : technology@barc.gov.in